| |
Voluntary (Comfort) Respirator Use of Respirators
Under some circumstances, employees may wish to use respiratory
protection equipment for their own comfort or sense of well being,
even when there is no recognized hazard or over exposure. In these
cases, not all of OSHA's respiratory protection requirements apply
(i.e., fit testing is not required). In order to voluntarily use
respiratory protective equipment in this way, the following criteria
must be met:
- There is no recognized hazard or potential for over
exposure
- The respirator must be NIOSH certified
- The respirator must be cleaned, stored, and maintained as
specified in Care of Respiratory Protective Equipment, below
- The respiratory protective equipment does not itself present a
hazard to the user
- The employee is medically cleared to use a respirator
(Exception: filtering facepieces, i.e., dust masks, "duck bill"
respirators, etc., do not require a medical clearance for
voluntary use)
- The employee is given a copy of the appendix Information for
Employees using Respirators when not required Under the Standard,
attached.
Employees who would like to voluntarily wear a respirator may
purchase and wear their own respirator or may ask their department to
supply one for them.
Medical Clearance
OSHA's revised respiratory protection program does not require
voluntary respirator users to undergo medical clearances. However, it
does require the employer to determine that the respirator itself is
not a hazard to the employee. To this end, EH&S strongly
recommends that voluntary respirator users seek a medical clearance
from University Health Service, Occupational Medicine, or their
personal physician before using a respirator. Medical clearances
should be repeated according to the physician's recommendation.
Cleaning and Disinfecting
The Occupational Health and Safety Administration have set
guidelines for the cleaning of respiratory protective equipment.
These are listed below. Alternatively, respiratory protective
equipment can be cleaned according to the manufacturers
recommendations so long as the equipment is cleaned and disinfected
in a way that does not damage it, and does not harm the user.
- Remove filters, cartridges, or canisters. Disassemble face
pieces by removing speaking diaphragms, demand and pressure-demand
valve assemblies, hoses, or any components recommended by the
manufacturer. Discard or repair any defective parts.
- Wash components in warm (43 deg. C [110 deg. F]
maximum) water with a mild detergent or with a cleaner recommended by
the manufacturer. A stiff bristle (not wire) brush may be used to
facilitate the removal of dirt.
- Rinse components thoroughly in clean, warm (43 deg. C [110
deg. F] maximum), preferably running water. Drain.
- When the cleaner used does not contain a disinfecting agent,
respirator components should be immersed for two minutes in one of
the following:
- Hypochlorite solution (50 ppm of chlorine) made by
adding approximately one milliliter of laundry bleach to one liter of
water at 43 deg. C (110 deg. F); or,
- Aqueous solution of iodine
(50 ppm iodine) made by adding approximately 0.8 milliliters of
tincture of iodine (6-8 grams ammonium and/or potassium iodide/100 cc
of 45% alcohol) to one liter of water at 43 deg. C (110 deg. F); or,
- Other commercially available cleansers of equivalent disinfectant
quality when used as directed, if their use is recommended or
approved by the respirator manufacturer.
- Rinse components thoroughly in clean, warm (43 deg. C [110
deg. F] maximum), preferably running water. Drain. The importance
of thorough rinsing cannot be overemphasized. Detergents or
disinfectants that dry on face pieces may result in dermatitis. In
addition, some disinfectants may cause deterioration of rubber or
corrosion of metal parts if not completely removed.
- Components should be hand-dried with a clean lint-free cloth or
air-dried.
- Reassemble face piece, replacing filters, cartridges, and
canisters where necessary.
- Test the respirator to ensure that all components work
properly.
Storage
Respiratory protection equipment must be stored in a way that
protects them from damage, dust, contamination, sunlight, chemicals,
excessive moisture, and extreme temperatures. They must also be
stored in such a way that it is unlikely that the face piece or
valves will be damaged or deformed. Additionally, emergency
respirators must be stored in an easily accessible way in the
workplace, must be in containers clearly labeled as containing
emergency respirators, and according to any manufacturers
recommendations. Self-contained breathing apparatus (SCBA) cylinders
must be stored fully charged, and must be recharged whenever they
fall below 90% full.
Inspection
The following checks are required as part of the respirator
inspection procedure:
- Respirator function
- Tightness of connections
- Condition of the face piece, head straps, valves, connecting
tubes, and cartridges, canisters, or other filters
- Pliability of any elastomeric parts
- Signs of cracking, discoloration, or other symptoms of
aging
- Tank pressure (SCBAs)
- Regulator and pressure alarm bell function (SCBAs)
- Tank condition (SCBAs)
When using respirators routinely, these inspections must be
performed before each use and during each cleaning. SCBAs must be
checked at least monthly. Inspection records must be kept until the
time of the next inspection.
Emergency respirators must also be inspected by performing the
checks above. These checks must be certified by tagging or labeling
the respirator with the date of the inspection, the name and/or
signature of the inspector, a serial number or other identifying
means, the condition of the respirator, and any needed repairs or
other maintenance. Emergency respirators shall be checked before
being brought into the workplace, and periodically thereafter.
Repairs
Respirators found to be defective or in need of repairs must be
removed from service immediately. When repairing a respirator or
replacing cartridges, valves or other components, only approved parts
shall be used to keep the NIOSH approval valid. No attempts, under
any circumstances, should be made to change, modify, or improve any
respiratory protection device. Only specially trained and qualified
technicians shall make repairs to SCBA's, pressure regulators and
other sensitive devices.
Training and Assistance
EH&S will provide training or other assistance to voluntary
respirator users upon request. Contact EH&S at x5-3241 for more
information.
Information for Employees Voluntarily Using Respirators When
Not Required Under the Standard
Respirators are an effective method of protection against
designated hazards when properly selected and worn. Respirator use is
encouraged, even when exposures are below the exposure limit, to
provide an additional level of comfort and protection for workers.
However, if a respirator is used improperly or not kept clean, the
respirator itself can become a hazard to the worker. Sometimes,
workers may wear respirators to avoid exposures to hazards, even if
the amount of hazardous substance does not exceed the limits set by
OSHA standards. If your employer provides respirators for your
voluntary use, of if you provide your own respirator, you need to
take certain precautions to be sure that the respirator itself does
not present a hazard. You should do the following:
- Read and heed all instructions provided by the manufacturer on
use, maintenance, cleaning and care, and warnings regarding the
respirators limitations.
- Choose respirators certified for use to protect against the
contaminant of concern. NIOSH, the National Institute for
Occupational Safety and Health of the U.S. Department of Health and
Human Services, certifies respirators. A label or statement of
certification should appear on the respirator or respirator
packaging. It will tell you what the respirator is designed for and
how much it will protect you.
- Do not wear your respirator into atmospheres containing
contaminants for which your respirator is not designed to protect
against. For example, a respirator designed to filter dust particles
will not protect you against gases, vapors, or very small solid
particles of fumes or smoke.
- Keep track of your respirator so that you do not mistakenly use
someone else's respirator.
If there are any questions or problems regarding this page this page, please contact Ken Simolo at
|